Series: RHP: RH/Temperature Sensor Transmitter; Accuracy: 3% Accuracy; Housing Type: Duct Mount w/Plastic Tip; RH Output: 0 to 10V; Temperature Sensor/Output: 0 to 10V Solid State.
Temperature Range: -40 to 140°F/(-40 to 60°C); Relative Humidity Range: 0 to 100% RH;
Hysteresis: ±1%; Repeatability: ±0.1% typical; Temperature Limits: -40 to 140°F/(-40 to 60°C); Storage Temperature: -40 to 176°F/(-40 to 80°C); Compensated Temperature Range: -4 to 140°F/(-20 to 60°C).
Power Requirements: 10 to 35V DC; Output Signal: 4 to 20 mA, 2 channels for humidity/solid state temperature, sensor models (loop powered on RH); Response Time: 15 seconds; Electrical Connections: Removable screw terminal block; Conduit Connection: Duct mount: 1/2˝ NPS; OSA: 1/2˝ (22.3 mm); Drift: <1% RH/year
RH Sensor: Capacitance polymer; Housing Material: Duct Mount: PBT; OSA: Polycarbonate; Enclosure Rating: NEMA 4X (IP65) for OSA mount only